Accreditation

For all CCRMN spectrometers :

- An accreditation is mandatory for all spectrometers:

For all Spectrometre

- Neither space nor special caracters including + in the file names. One can only use letters, numbers, _ and -. Dot is only permitted once and shall be followed by any latter caracter.

- The calibration of the spinners on the samples must be done carefully , and the tube must not float in the spinner (risk of a de-calibration).

For AV300 routine experiments

- It is forbidden to borrow the spinners from the AVL300 sample changer to run your samples on the self-service Av300.

- Put the spectrometer back into automatic mode before disconnecting.

- Put the fictional spinner in the spectrometer if there is nobody after you.

For continuous sample changer

- Do not use broken, too short, or thickly parafilmed tubes, or Young tubes that may not be properly picked up by the sample changer clamp.

- Respect the color code of the caps by laboratory.

- Before taking your sample back, wait until the clamp is no longer under it (except if there is no analysis on any tube left).

Topspin for zero :

  • video tutorial [ See more ]